Ready For Sale
Secondhand Alpinion Ecube 7 Ultrasound Machine
Price: USD$ 7.000,00 Approx: 315.000,00 TL
Ready For Sale
Ask a Question
Payment
No additional fees, full assurance. We provide complete financial and operational security in secondhand medical device trading. For this, we offer the "Secure Payment" service. This free service protects the rights of both parties by securing the buyer's money and the seller's product. The Secure Payment system is a standard assurance mechanism offered by Medbidding. For additional information, review the "services" page.
There is no cash on delivery order system on the Medbidding platform. For payments to be made by credit card, the product to be purchased must comply with this payment method. You can contact us to get information about this. We would be happy to assist you.
For payments made outside of Turkiye, you can choose bank transfer, credit card, Western Union or cryptocurrency options. Installment options are not currently available for credit cards other than Turkish banks.
Shipping
Standard Shipping Conditions
In order to ensure secure transactions on Medbidding, the shipping process is managed through four different scenarios depending on the location of the buyer and seller. Free shipping is available for some categories. The terms below apply to all categories unless otherwise stated.
Buyer and Seller in Turkiye
- Seller → Medbidding: The seller packages the product and sends it to the Medbidding operation center. The seller is responsible for this shipping cost.
- Medbidding → Buyer: The product, inspected by Medbidding engineers, is sent to the buyer's address. The buyer is responsible for this shipping cost.
Buyer Outside Turkiye, Seller in Turkiye
- Seller → Medbidding: The seller packages the product and sends it to the Medbidding operation center. The seller is responsible for this shipping cost.
- Medbidding → Buyer: The product, inspected by Medbidding engineers, is sent to the buyer's address. The buyer is responsible for this shipping cost.
Buyer in Turkiye, Seller Outside Turkiye
- Seller → Medbidding: The seller packages the product and sends it to the Medbidding operation center. The seller is responsible for this shipping cost.
- Medbidding → Buyer: The product, inspected by Medbidding engineers, is sent to the buyer's address. The buyer is responsible for this shipping cost.
Buyer and Seller Outside Turkiye
If there is a local operation center in the seller's country:
- Seller → Medbidding: The seller packages the product and sends it to the Medbidding operation center. The seller is responsible for this shipping cost.
- Medbidding → Buyer: The product, inspected by Medbidding engineers, is sent to the buyer's address. The buyer is responsible for this shipping cost.
If there is no operation center in the seller's country:
- Technical Inspection: Inspections are performed by our engineers via remote video call.
- Seller → Buyer: The seller packages the product and sends it directly to the buyer's address. The seller is responsible for this shipping cost.
Objective AI Report
Disclaimer: I am Medbidding AI. I am an unbiased AI robot. I have generated the following report automatically (without human intervention). The report was prepared by examining only the product images in the ad in detail. The report may contain errors. Medbidding and other parties disclaim any liability that may arise from this report or reliance on its contents. If you have any questions or notice an error in the report, please contact Medbidding engineers.
Report date: 10.02.2026
ALPINION E-CUBE 7 Ultrasonography Device Analysis Report
Device Identification and Brand/Model Analysis
As a result of a detailed visual inspection, the product displayed has been identified as a medical ultrasonography (ultrasound) device. The device is a system mounted on a mobile station, consisting of a monitor, control panel, and main processing unit.
The following details have been definitively determined during the brand and model inspection of the product:
- Brand: ALPINION (Clearly written on the lower bezel of the device monitor and potentially on the device’s startup/standby screen.)
- Model: E-CUBE 7 (Clearly indicated with a logo on the handgrip at the bottom of the control panel and on the front and side surfaces of the device’s main casing.)
The device is not a patient bed or patient transport lift; therefore, additional component inspections specific to these products (motor, battery brand, etc.) are not included in this report.
General Condition and Physical State
Upon inspecting the device’s external cosmetic condition, it is observed that despite being second-hand, its overall structural integrity is maintained.
- Casing Structure: No deep cracks or fractures have been detected in the gray and anthracite plastic coatings forming the main body. However, slight surface dust accumulation due to use may be visible.
- Physical Deformations: There are normal signs of use on the base of the device’s support stand and on the wheel covers. No significant dents are noticeable on the side covers.
- Cleanliness: While the device generally appears clean, there is a high probability of dust accumulation between cables and in probe holders. It may require professional surface cleaning.
Control Panel and Mechanical Components
The control panel, which is the device’s user interface, appears complete.
- Keypad and Buttons: The QWERTY keyboard, TGC (Time Gain Compensation) sliders, rotatable mode dials, and function keys are physically in place. No erasure or missing keycaps have been detected on the keys in the visual inspection.
- Trackball: The trackball located in the center of the panel is present and appears sturdy along with the surrounding “Set/Exit” buttons.
- Monitor Mechanism: The monitor is connected to the device’s body via a movable arm mechanism. This mechanism maintains its upright position; no sagging has been observed.
- Wheels: The device’s mobile stand features wheels that are presumed to be lockable.
Electronic Components and Screen Analysis
The device is in the off position during the inspection.
- Screen: A flat panel monitor with LCD/LED technology is present. Other than light reflection, no significant deep scratches or pixel fractures are visible on the screen surface. As the device is off, the screen’s image quality, software error messages, or pixel status could not be analyzed.
- Connection Sockets: The device’s rear panel features various ports for video and data transfer (S-Video-like inputs, audio outputs). Additionally, USB input ports are present on the side/bottom of the control panel.
Accessories and Probe Information
The probes, which are the most critical components of an ultrasound device, have been visually inspected.
- Number of Probes: 2 units of probes are physically connected to the device.
- Probe Condition:
- Probe cables are hung on hooks. No serious crushing or breakage of the cables was detected visually, but their winding appears somewhat tangled. Slight age-related yellowing may be present in the cable insulation colors (cream/white).
- Probe heads (transducer parts) are in their holders. Their frequency or type (Convex, Linear, etc.) could not be technically confirmed as the writings on them are too small in the visual, but based on their visual form, they can be said to be standard medical imaging probes.
- Missing Items: The device’s printer slot is empty, or no thermal printer accessory is visible in the image.
Rear Panel, Labels, and Connections
The rear part of the device (Image 4) contains technical details:
- Labels: On the rear panel, there are yellow “Caution / Attention” warning labels. These labels typically include warnings for laser safety, electric shock risk, or transportation.
- Serial Number/Lot Information: On the lower rear part, there is an area on a gray label containing technical information (input voltage, model number, etc.), but the visual resolution does not allow these writings (Serial No, Ref Code) to be read.
- Power Input: At the bottom, a standard 3-pin power input (Power Socket) and a red main power switch (I/O) are visible next to it. A connection screw for a grounding cable is also present.
- Cooling: The grilles on the rear body belong to the ventilation system and are open; no blockage is visible.
Areas of Use and Purpose
ALPINION E-CUBE 7 is a multi-purpose ultrasonography system. It is suitable for use in Radiology, Obstetrics/Gynecology, Cardiology, Urology, and general internal medicine imaging. Depending on the type of probes in the visual, imaging of abdominal organs, pregnancy follow-up, or superficial tissue examinations can be performed. It is a suitable and portable model for hospitals, clinics, and imaging centers, thanks to its wheeled structure, allowing it to be moved within rooms or between departments.
Potential Malfunction Risk and Evaluation
Based on the visual analysis, the device’s body condition can be described as “Good”.
- Probe Cables: The tangled arrangement of the cables may create a potential risk of internal wire breakage; signal tests are recommended before use.
- Contact Points: Given the age of the device, there is a mechanical possibility of intermittent contact occurring over time in the “Gain” knobs or TGC sliders on the control panel due to dust accumulation, but there is no visual damage currently.
- Software/Hardware: As the screen is off, no comments can be made regarding the hard disk status, operating system health, or crystal defects.
Summary: The product is a physically clean, complete keypad, ALPINION E-CUBE 7 brand/model medical ultrasound device with 2 probes.